360203-1/B
11
Installation & Configuration
6.6
Connect Backup Power
NOTE
►
The standard backup battery system provides electrical power to the i.C
3
monitoring system, Access Control magnetic
door lock and compartmental access communication boards. It may also be referred to as the i.C
3
Monitoring System /
Access Control backup battery.
The i.C
3
Monitoring System and Access Control magnetic door lock have a standard backup battery system,
enabling a period of continuous operation if power is lost. The i.C
3
Monitoring System / Access Control backup
battery ON/OFF switch is located behind the access panel in the lower right front of the Compartmental Access
Refrigerator.
i.C
3
Monitoring System / Access Control backup battery ON/OFF switch (circled)
The i.C
3
Monitoring System / Access Control backup battery is switched
OFF
for shipping. Switch the backup
battery ON/OFF switch
ON
to provide backup power to the i.C
3
monitoring system and Access Control magnetic
door lock.
NOTE
The i.C
3
Monitoring System and Access Control magnetic door lock will start on battery power alone. If the Compartmental Access
Refrigerator is not connected to AC power and the battery is switched on, the i.C
3
Monitoring System and Access Control magnetic
door lock will begin running on battery power.
Battery life varies by manufacturer as well as voltage level remaining. If full battery power is available and no
battery-related alarms are active, backup power for the i.C
3
Monitoring System and Access Control magnetic door
lock is available for up to 20 minutes.
The Low Battery alarm will activate when battery power is almost depleted.
During a power failure:
►
The standard backup battery system does not provide continued refrigeration of the chamber. The chamber temperature
may rise above the established limits necessary to maintain integrity of stored product.
►
The standard backup battery system provides power to the monitoring system and Access Control magnetic lock until battery
power is depleted.
►
The standard backup battery system is rapidly depleted by the Access Control magnetic lock.
NOTE
If AC power is lost, the monitoring system will automatically disable some features to prolong battery power. Data collection will
continue until battery power is depleted.