Prepare For Monitoring
7.1
Temperature Probes
Temperature probes are fragile; handle with care.
NOTE
Remote probes may also be introduced through the back panel.
Four types of temperature probes are included on the Compartmental Access Refrigerator, air probe, primary
probe, condenser probe and control probe.
Air probe
Primary probe (with probe
Condenser probe
Control probe
bottle and bottle holder)
7.1.1 Air Probe
The air probe is located along the back right side of the unit.
7.1.2 Primary Probe
The primary probe is located behind the grill in the top right side of the chamber. To access the probe, the grill
must be removed.
For each probe bottle, use:
►
Approximately 4 oz. (120 mL) of product simulation solution (10:1 ratio of water to glycerin)
Fill Temperature Probe Bottle
1
Remove all probes from bottle and remove bottle from bracket.
2
Remove cap and fill with approximately 4 oz. (120 mL) of product simulation solution.
3
Secure cap on bottle and place in bracket.
4
Replace probes, immersing at least 2” (50 mm) in solution.
7.1.3 Condenser Probe
The condenser probe is attached to the condenser discharge line.
7.1.4 Control Probe
The control probe is located within the evaporator housing.