Event Log Codes
Event Code
Description
Event Code
Description
4
Communication Failure 1
29
Agitator 1 Low Speed
5
Communication Failure 2
32
Agitator 1 Maintenance Required
6
Communication Failure 3
35
Condensate Tray Full
7
Control Probe Failure
36
Drive Space Low
8
Primary Monitor Probe Failure
37
Drive Space Full
11
Door Open
38
Heat Pump 1 Connection Error
12
Power Failure
41
Heat Pump 1 Failure
13
Date/Time Change
43
Heat Pump 1 High Temp
14
High Temperature Alarm Test
47
MPB Communication Failure
15
Low Temperature Alarm Test
48
Primary Monitor Probe High Temperature
18
No Battery
49
Primary Monitor Probe Low Temperature
19
Low Battery
56
External Fan 1 Failure
20
Power-Up
57
External Fan 2 Failure
23
Agitator 1 Communication Failure
60
Internal Fan 1 Failure
26
Agitator 1 High Speed
1. Communication Failure 1
• Triggered if communication is lost between i.C
3
display board and control board
•
Unit will continue to run with previously saved settings
•
Screen will not display temperature changes or alarm conditions
• i.C
3
system will continue to reset until connection is re-established
• No data is logged until the alarm is cleared
2. Communication Failure 2
• Triggered if communication is lost between i.C
3
display board and internal system memory
•
Unit will continue to run with previously saved settings
3. Communication Failure 3
•
Triggered if the database is corrupted
• The database is archived and a new database is automatically created
•
Unit will continue to run with previously saved settings
Note
The "Communication Failure 3" alarm indicator appears on the Home screen until the event has been acknowledged.
Helmer Scientific
i.C
3
User Guide for i.Series Incubators
360406/A
46