Alarm Settings
>
>
Alarm Settings
Setting
Description
Default Value
Range
Primary Monitor Probe High Temp (setpoint)
High temperature at which alarm condition occurs
24 °C
-100 - +50 °C
-148 - +122 °F
Primary Monitor Probe High Temp (time delay) Time after high temperature alarm condition occurs
until alarm sounds
0 minutes
0 - 20
Primary Monitor Probe Low Temp (setpoint)
Low temperature at which alarm condition occurs
20 °C
-100 - +50 °C
-148 - +122 °F
Primary Monitor Probe Low Temp (time delay) Time after low temperature alarm condition occurs
until alarm sounds
0 minutes
0 - 20
Heat Pump 1 High Temp (setpoint)
High temperature at which alarm condition occurs
50 °C
Heat Pump 1 High Temp (time delay)
Time after high temperature alarm condition occurs
until alarm sounds
1 minute
0 - 99
Power Failure
Time after power failure occurs until alarm sounds
1 minute
0 - 99
Probe Failure
Time after probe failure occurs until alarm sounds
0 minutes
0 - 99
Door Open
Time door remains open until alarm sounds
1 minute
0 - 99
Agitator 1 Hi Speed Alarm
High speed at which alarm condition occurs
75
21 - 90
Agitator 1 Low Speed Alarm
Low speed at which alarm condition occurs
65
20 - 89
Agitator 1 Motion Alarm Delay
Time after speed alarm condition occurs until
alarm sounds
1 minute
0 - 99