Agitator Status

The Agitator Status screen provides real time information regarding the agitator speed and cycle count.

 View Agitator Speed History Graph

Touch the button showing the current agitator speed. The 

Agitator Speed History graph will appear.

 Reset Cycle Count

1.  Touch the 

 to reset the cycle count. The 

"Are you sure you want to reset the count for Agitator X?"

 message appears.

2.  Touch the 

 to confirm or the 

 to cancel.

Information Found in Agitator Status

•  Serial number
• 

Speed (cycles per minute within the last 60 minutes)

• 

Maximum speed (within the last 60 minutes)

• 

Minimum speed (within the last 60 minutes)

• 

Door openings (represented by a yellow vertical line when the show/hide button is set to show)

Helmer Scientific

i.C

 User Guide for i.Series Incubators

360406/A

16

Содержание i.C3

Страница 1: ...i C User Guide For Helmer i Series Incubators 360406 A Countertop PC100 Pro PC900 Pro PC1200 Pro...

Страница 2: ...is installed Confidential Proprietary Notices Use of any portion s of this document to copy translate disassemble or decompile or create or attempt to create by reverse engineering or otherwise the in...

Страница 3: ...11 7 Temperature Graph 12 Event Data 13 Viewable with the 1 Day Time Span 13 8 AgiTrak if installed 15 Agitator Status 16 Agitator Count Log 17 Agitator Diagnostics 17 9 Information Logs 18 Event Log...

Страница 4: ...Systems 35 Restore Factory Settings 35 System Functions 36 13 Access Control if installed 37 Access Control Setup 37 Access Control Screen 38 Access Control as the Home Screen 39 14 Icon Transfer 40 A...

Страница 5: ...strument bezel as are additional features Touchscreen The interface to the i C3 system USB Port Connect a flash memory device to download temperature data or upload firmware Ethernet RJ45 Connection N...

Страница 6: ...dicators and Buttons For a complete list of all icons and indicators see the Icon Reference Guide at the end of this user guide Application icons serve as navigation buttons to the associated applicat...

Страница 7: ...ed The selected language is displayed in the Language button 3 Touch to confirm language selection or to cancel 4 Touch the Home icon to continue to the Home screen Note English is the default languag...

Страница 8: ...ence within a given period of time Application icons Icons located at the bottom of the screen which include five factory preset application icons From left to right they are i C3 Applications APPS Se...

Страница 9: ...View product support information including the firmware revision numbers i C3 Settings All i C3 system settings are configured and other preferences selected from the i C3 Settings screen Note The i...

Страница 10: ...tion log s PDF Downloads Download historical information as a PDF Uploads Upload firmware updates Alarm Test Initiate automatic alarm tests for the high and low temperature alarms AgiTrak Control agit...

Страница 11: ...rature changes are displayed in the Temperature Display on the left side of the screen As the temperature moves into alarm conditions the temperature reading turns red The message Peltier Test Probe C...

Страница 12: ...e time reaches the maximum display of 999 hours and 60 minutes the message will display 999 60 but minimum and maximum temperatures will continue to be tracked Reset Min Max temperature 1 Touch the Re...

Страница 13: ...alarm limits Graph is displayed in strip chart form A black dashed vertical line appears on the graph where there is a change in the date or time A white dashed vertical line appears on the graph wher...

Страница 14: ...ent data 1 Touch the Alarm Condition icon to navigate to the Event Log screen 2 Touch the Event to view detailed event data Graph Display of Alarm Test Alarm tests are shown on the graph by a small Al...

Страница 15: ...point on the graph Using one touch 1 Touch the desired point on the graph A dialog box will appear providing selected date time and temperature data Note The one touch feature should be used for refer...

Страница 16: ...int spin box to select desired agitator speed 3 Select the Agitator Serial Number button A numeric keyboard will appear 4 Enter the access code 3773 and touch the to confirm A numeric keyboard will ap...

Страница 17: ...eset the cycle count The Are you sure you want to reset the count for Agitator X message appears 2 Touch the to confirm or the to cancel Information Found in Agitator Status Serial number Speed cycles...

Страница 18: ...Number Count at Reset Agitator Diagnostics The Agitator Diagnostics screen provides information regarding the operational status of the installed agitator Notes Use the current Administrator level pa...

Страница 19: ...most recent events can be viewed on the Event Log screen Refer to Data Transfer Downloading Temperature and Event Data for more information Viewing the event log 1 Touch and hold the bidirectional ar...

Страница 20: ...Log Formatting Bold text Not yet viewed Normal text Viewed Dimmed text Downloaded Data row shaded blue High or low alarm triggered from an alarm test Event Log Detail Screen The Event Log Detail prov...

Страница 21: ...g screen The date and time of the event acknowledgement is automatically added to the Signature button and the event information can no longer be edited or touch the to cancel the Save operation Note...

Страница 22: ...Override Date of access Time of access Length of time door remained open Method of access PIN Key or Remote Access Log Formatting Bold text New data since the last time the log was viewed Normal text...

Страница 23: ...lash Drive clears and the Data Selection spin buttons are no longer shaded 2 Touch the Date Range buttons to select the desired beginning left and ending right dates The Data Report Calendar screen is...

Страница 24: ...iliary Systems for information regarding enabling the USB port The Download screen can not be closed during a data download unless an error occurs Up to 3 months of data may be downloaded at a time If...

Страница 25: ...on to begin the download The Download Status bar becomes active showing the progress of the download When finished the message Data download complete along with the number of files downloaded A date a...

Страница 26: ...value These values will be referenced if the firmware update causes a change to the offset values The Upload screen can not be closed during a data upload unless an error occurs If the USB is disabled...

Страница 27: ...or control processor CP firmware version number have been updated If either firmware version number has not been updated from the recorded version number contact Helmer Technical Service For more deta...

Страница 28: ...4 hour temperature graph The screensaver is automatically displayed after two minutes of inactivity on the Home screen Touch anywhere on the screensaver to return to the Home screen Notes Inactivity o...

Страница 29: ...gh the settings menu A password protects i C3 settings from unauthorized changes Note Helmer recommends password protection remain ON Accessing Settings If password protection is turned ON touch the i...

Страница 30: ...etpoint Alarm Settings Parameters for activating an alarm may be set through Alarm Settings Set alarm parameters 1 Touch the Alarm Settings button 2 Touch the minus or plus on the corresponding Setpoi...

Страница 31: ...e password 1 Touch the Users and Password button 2 Touch the Change Password button a numeric keypad is displayed 3 Enter a new 4 digit password and touch 4 Re enter the 4 digit password to confirm 5...

Страница 32: ...desired setting Set alarm volume 1 Touch the Sounds Settings button 2 Touch the minus or plus on the Alarm Volume spin box to select desired volume Select alarm tone 1 Touch the Sounds Settings butto...

Страница 33: ...button to select the desired format Enable disable Temperature Graph Screensaver 1 Touch the Display Settings button 2 Toggle the Temperature Graph Screensaver button On or Off to select desired sett...

Страница 34: ...Device Status Current sensor readings and device status are displayed on the Device Status screen Helmer Scientific i C3 User Guide for i Series Incubators 360406 A 33...

Страница 35: ...acent spin box until the correct value is displayed Adjust temperature setpoint 1 Touch the Sensor Calibration button 2 Touch the minus or plus on the corresponding probe spin box to adjust the offset...

Страница 36: ...is interface Restore Factory Settings Settings will be restored to their factory default through Restore Factory Settings Restore factory Settings 1 Touch the Restore Factory Settings button The Resto...

Страница 37: ...system 1 Touch the System Functions button 2 Touch the Shut down button The confirmation box appears 3 Touch to confirm or to cancel Notes The i C3 system should be shut down prior to shutting down t...

Страница 38: ...e Supervisor PIN must be used to set up user profiles The Supervisor PIN does not allow access to the unit At least one user ID must be set up to gain access to the unit The Supervisor PIN should be c...

Страница 39: ...3 Enter a new 4 digit PIN 4 Touch to confirm The numeric keypad closes and the new PIN is displayed in the table or touch to cancel Access Control Screen Temperature and Alarm Conditions are displayed...

Страница 40: ...e screen is disabled Refer to Icon Transfer for more information If the Access Control Home Screen is disabled the keypad is accessed by touching the Access Control icon Enable disable Access Control...

Страница 41: ...yed on the Home screen Reposition icons Drag icons with a touch drag motion of the finger Drag the selected icon directly over the icon that is currently located in the targeted position Note The i C3...

Страница 42: ...t To display the lowercase extended character keyboard touch the Lowercase button Touch and hold the Back button to delete multiple characters i C3 Apps Buttons Alphanumeric Keyboard Buttons Toggle bo...

Страница 43: ...Navigate to the i C APPS screen Alarm Conditions icon indicator Indicates alarm event occurred Navigate to the Event Log i C Settings Navigate to the i C Settings screen Cancel Test Touch to cancel A...

Страница 44: ...Set at the factory Temperature Units Temperature units in C or F C C or F Temperature Graph Screensaver Display temperature graph as Home screen On Alternate Screensaver optional Display alternate scr...

Страница 45: ...ge Agitator Selection if enabled Select agitator location within incubator 1 1 3 Agitator 1 Speed Setpoint if enabled Set agitator speed cycles per minute CPM 72 40 80 Helmer Scientific i C3 User Guid...

Страница 46: ...t Pump 1 High Temp setpoint High temperature at which alarm condition occurs 50 C Heat Pump 1 High Temp time delay Time after high temperature alarm condition occurs until alarm sounds 1 minute 0 99 P...

Страница 47: ...lure 60 Internal Fan 1 Failure 26 Agitator 1 High Speed 1 Communication Failure 1 Triggered if communication is lost between i C3 display board and control board Unit will continue to run with previou...

Страница 48: ...Helmer Scientific 14400 Bergen Boulevard Noblesville IN 46060 USA PH 1 317 773 9073 FX 1 317 773 9082 Copyright 2019 Helmer Inc 360406 A...

Отзывы: