2.
Touch the Upload icon to begin data transfer. The upload progress bar is displayed.
3. When finished, the
"Data Upload Completed"
message and the date and time are displayed. The screen then refreshes
to display the
"Wait"
screen, and the system automatically powers down and restarts.
4.
Remove the memory device from the USB port. Updated firmware is loaded to the i.C
3
memory.
5. After the i.C
3
system restarts, verify the display processor (DP) and/or control processor (CP) firmware version number
have been updated. If either firmware version number has not been updated from the recorded version number, contact
Helmer Technical Service. For more details regarding referencing the firmware revision number, refer to
i .C
3
Applications (APPS), Using i .C
3
Applications
6.
Verify the Control Sensor Offset temperature value has not changed. If the temperature offset value has been changed
from the recorded value, calibrate the offset value. For details on calibrating the offset values, refer to
i.C3 Settings, Sensor Calibration.
Helmer Scientific
i.C
3
User Guide for i.Series Incubators
360406/A
26