Uploading Firmware Updates
Use the Upload screen to upload firmware updates. If the Upload icon is touched before a memory device is inserted,
the message
"Insert USB Flash Drive"
is displayed until a memory device is inserted.
Notes
•
Record the display processor and control processor firmware version number before uploading a firmware update.
These numbers will be referenced after the update to ensure the update was successful.
•
Record the Control Sensor Offset temperature value. These values will be referenced if the firmware update causes a
change to the offset values.
•
The Upload screen can not be closed during a data upload unless an error occurs.
•
If the USB is disabled in the Auxiliary Setting screen, the message "USB port is currently disabled" is displayed. Refer to
i .C
3
Settings
,
Auxiliary Systems
for information regarding enabling the USB port.
Upload firmware
1.
Insert a memory device into the USB port located on the right of the i.C
3
monitor bezel. When the memory device is
detected, the message
"Insert USB Flash Drive"
clears and the Upload icon is no longer shaded.
Helmer Scientific
i.C
3
User Guide for i.Series Incubators
360406/A
25