Sensor Calibration
Calibration adjusts temperature setpoints so the value displayed matches the actual chamber temperature, as
measured by an independent thermometer. View temperature probe readings and adjust offsets through Sensor
Calibration.
Sensor Reference
Indicates probe location on the control board.
Current Reading (+ Offset)
View the real-time temperature reading which includes offset calculation.
Offset
The spin boxes are used to calibrate the temperature probe readings using the offset values. Touch the minus (-) or
plus (+) on the adjacent spin box until the correct value is displayed.
Adjust temperature setpoint
1. Touch the
Sensor Calibration
button.
2.
Touch the minus (-) or plus (+) on the corresponding
probe
spin box to adjust the offset.
Note
Primary probe temperature offset may require periodic calibration. Refer to the product service manual for instructions
regarding calibrating a chamber temperature probe.
Date/Time
Date and time values can be changed through the Date/Time
screen.
Change/set date and time
1. Touch the
Date/Time
button.
2.
Touch the minus (-) or plus (+) on the corresponding
Date
or
Time
spin box to select the appropriate value.
3.
Toggle the
AM/PM
button to select AM or PM.
4. Touch
to save.