6 Min/Max Temperature Monitoring
The Min/Max temperature display shows the highest and lowest Primary Monitor probe temperature occurrence since
the last system reset (power-on event) or manually-initiated reset.
Notes
•
The Min/Max temperature display can be turned on or off. Refer to
Display Settings
for more information.
•
Once the time reaches the maximum display of 999 hours and 60 minutes, the message will display ">999:60", but
minimum and maximum temperatures will continue to be tracked.
Reset Min/Max temperature
1. Touch the
Reset
icon. The message "Reset min/max temperature memory?" is displayed along with the time (in hours
and minutes) since the last reset.
2. Touch
to confirm. The timer is reset to 000.00 and the minimum and maximum recorded temperature is cleared.
Or, Touch
to cancel.
Helmer Scientific
i.C
3
User Guide for i.Series Incubators
360406/A