Calibration
WTX120
A4500-1.1
HBM: public
145
11.12
W&M Info
For more information on the keys necessary to access this menu group, go to
the
In the
W&M Info
group you can check that the set scale parameters match the
values stipulated for scales subject to weights and measures legislation. Any
deviations are displayed.
Menu
Key
Function/Secondary selection
Select Group 1-9
Press 9 times for group
9 W&M Info
W&M Setup ok
Display when setting is correct
Error: Overload > 9d
Display when the set values do not com
ply with approval requirements
Example: Value for overload display sup
pression is set higher than 9d.
Continue
→ Check the settings (allowable values in brackets)
Motion Window
(0,5d)
Motion Counter
(
≥
7)
Autozero Range
(0,5d)
Pushbutton Range
(
≤
4%)
Overload
(max. 9d)