SupervisorMode
106
A4500-1.1
HBM: public
WTX120
When the filling process starts (RUN command or digital input or Enter key) the
tare delay starts. Taring is then carried out when the tare delay ends.
Requirements
In upward dosing, the empty weight and coarse flow cut-off point must not be
exceeded on starting.
Tare mode must be active (
Tare Mode On
).
Activation
Activate taring after a delay by way of the tare delay, specifying the delay time
to apply (see also chapter
.
10.2.27
Start fine flow before coarse flow (Start Fine(s))
Set or read the duration of an initial fine flow phase before the coarse flow.
Information
The function is not active for downward weighing.
After starting or taring, and before the coarse flow, the fine flow signal is acti
vated for the preset time. You can use this additional fine flow time before the
coarse flow to prevent excessive foaming of the liquid being filled as a result of
the coarse flow.