Calibration
126
A4500-1.1
HBM: public
WTX120
→ Single-range scale
Three Intervals T+
, triple-range scale
Triple Range
or
multi-interval scale
Three Intervals
High Capacity
999999
Enter the maximum load of the highest range.
Example:
High Capacity 6000
High Int.
999.9999
Enter the interval of the highest range.
Example:
High Int. 1.0000
The error message
Entry Not Valid!
is dis
played if you enter an invalid scale interval.
Mid Capacity
999999
Enter the maximum load of the middle range.
Example:
Mid Capacity 3000
If the
Mid Capacity
value is greater than
High
Capacity
, the error message
Entry Not Valid!
is displayed.
Mid Int.
999.9999
Enter the interval of the middle range.
Example:
Mid Int. 1.0000
The error message
Entry Not Valid!
is dis
played if you enter an invalid scale interval.
Low Capacity
999999
Enter the maximum load of the lowest range.
Example:
Low Capacity 1500
If the
Low Capacity
value is greater than
Mid
Capacity
, the error message
Entry Not Valid!
is displayed.
Low Int.
999.9999
Enter the interval of the lowest range.
Example:
Low Int. 1.0000
The error message
Entry Not Valid!
is dis
played if you enter an invalid scale interval.
Содержание WTX120
Страница 1: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Страница 3: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Страница 184: ...Service password 182 A4500 1 1 HBM public WTX120 ...
Страница 185: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Страница 366: ...Service Passwort 182 A4500 1 1 HBM public WTX120 ...
Страница 367: ...Operating Manual Bedienungsanleitung Manuel d emploi English Deutsch Français WTX120 ...
Страница 549: ...Mot de passe de service WTX120 A4500 1 1 HBM public 183 ...