Clarification of important parameters
Group
Parameter
Meaning
DIALOG
Password
Specify (change) password, 0000 - 9999
(factory settings password: 0000)
PassStat
Specify password status:
active = password must be entered;
inactive = PME can be operated without entering a password
I.ParaS to
I.AddFnc
Access to group via keyboard enabled or disabled.
PARAM. SET
Load ?
You can either load the factory setting, or one of the four
parameter sets that are saved.
Save ?
Source
All the device settings can be saved power failsafe in four
parameter sets. Each time the mode of operation changes
from setting to measuring mode, you will be asked whether or
not the change is to be saved. The data are permanently
protected if you exit setting mode by confirming “Yes" to the
protection question.
Selection of signal values for combination
1)
:
Gross, Net, MinMax
MATHEM
CALCULATIO
Operation
Addition: adds the gross values from max. 8 input channels
(measurement modules)
NoOprnds
Number of modules configured (max. 8)
SlaveR.
PDO = module transmission rate
Group
Parameter
Meaning
CONDI
TIONING
Difference between taring/zero balance: The zero balance
(>0<) affects the gross and the net value. Taring (>T<) only
affects the net value.
>0<kN
2)
Enter zero value. Zeroing affects the gross and the net value.
>0<set ?
Trigger zero balance; set current measured value
(physical unit) to zero
>0<save
The zero value is transferred to the EEPROM every time
zeroing occurs (life: 100,000 cycles)
>T< kNm
1)
Enter tare value. Taring affects the net value.
>T<set ?
Trigger taring; net value becomes 0
>T<save
Save tare value immediately after taring