Control inputs and control outputs are available for the
MP70DPS7
and
MP70S7
versions:
Inputs /outputs
Plug terminal 3:
Here there are
4inputs
available to you for controlling the
functions of the PME.
Plug terminal 4:
Here there are
4 outputs
available.
Group
Parameter
Meaning
IN/OUT
1)
Outputs
1- 4
Outputs 1 - 4 can be assigned with the following functions for
each channel:
Limit values 1 to 4, standstill, error, inactive
Mode
Out1...4
The output signal is inverted (pos.log) or not inverted (neg.
log).
The listed functions can be freely assigned to the remote con
trols (inputs/outputs).
Functions
Input level 0 V
Input level 24 V
Taring
The transition from 0 V to 24 V starts the taring process
Zero
balance
The transition from 0 V to 24 V sets the current measurement
signal to zero
PkMomMax
Peak value operating mode
for PkMax
Current value operating mode
for PkMax
PkMomMin
Peak value operating mode
for PkMin
Current value operating mode
for PkMin
PkHldMax
PkMax and PkMin memory
contents are updated
PkMax and PkMin memory
contents are frozen
PkHldMin
PkMin memory contents are
updated
PkMax memory contents are
frozen
ParaCo1
Selecting parameter sets and binary
coded inputs
ParaCo2
Parameter set
ParaCo2
ParaCo1
1
0
0
2
0
1
3
1
0
4
1
1
1)
MP70DPS7 and MP70S7 only