61
PME-MP70DP
A0658-7.3 en
HBM
Limit1
Level at which limit switch 1 responds
2 or 4 bytes
Limit2
Level at which limit switch 2 responds
2 or 4 bytes
Limit3
Level at which limit switch 3 responds
2 or 4 bytes
Limit4
Level at which limit switch 4 responds
2 or 4 bytes
10.2.2 Configuration for MP01
Description
Meaning
Length
Channels
1 - 4
Gross
Gross measured value
2 or 4 bytes
Net
Net measured value (gross less tare value)
2 or 4 bytes
Max
The contents of the Maximum memory
2 or 4 bytes
Min
The contents of the Minimum memory
2 or 4 bytes
Pk‐Pk
Peak‐to‐Peak, difference between Max and Min
2 or 4 bytes
Status1
Status word with state of the limit value switches and
general error bits
2 bytes
Output values:
Description
Meaning
Length
Channels
1 - 4
Control
word
Control word for triggering taring, zeroing, clearing the
peak value store, parameter selection, etc.
2 bytes
Limit1
Level at which limit switch 1 responds
2 or 4 bytes
Limit2
Level at which limit switch 2 responds
2 or 4 bytes
The formats of the data content that is transmitted cyclically are specified in
detail in chapter 9.2.
The measured values are available either as 16‐bit integers or 32‐bit integers,
as required. The values are always scaled to physical quantities, with the
number of decimal places set in the module being used. The choice of
whether to use a 16‐bit (2 bytes) or 32‐bit format (4 bytes), is set in the
parameterization telegram.
Typical combinations are predefined in the GSE file. If you need different com
binations, you can use the specification from chapter 10.5.1 to extend the
GSE file accordingly.