Error displays
The permissible display range depends on the nominal value of the scale and
the set mode of operation (not legal for trade / OIML / NTEP).
MODE
³
LEGAL-FOR-TRADE
Lower display limit
Upper display limit
NO (not legal-for-trade)
−160 %
+160 %
OIML
−2 %
Nominal value + 9 d
NTEP
−2 %
Nominal value + 5 %
The percentages relate to the nominal weighing range (”NOMINAL VALUE”
parameter)
The following error message appears in the display when the measured value
is
outside the maximum display range: −−−−−−−−−−−−−−−−−
Other errors are displayed as a four-digit code with explanation
(e.g. ERROR!
CODE 5700
No standstill
No measured value acquired
They should not occur in normal operation (see also Chapter 12.2, page 36).
7.9
Standstill recognition
The zeroing, taring and printing functions are only executed if the value in the
display is stable. This is known as standstill and is indicated by showing the
unit of measurement. The condition for standstill is that the value changes by
no more than a specific fluctuation limit per time unit. With fluctuating (wind)
loads or a very high scale resolution, it is quite possible that standstill will ne-
ver be achieved. In this case, a more strongly damping filter or a lower resolu-
tion must be selected in the parameter setting.
The various options for standstill indication can be selected in the ”SCALE PA-
RAMETERS-PARAMETER
→
STANDSTILL MON.” menu (see system CD).
It is also possible to switch off the standstill conditions (but not for legal-for-
trade applications). The unit is always displayed when standstill monitoring is
switched off.