DIS2116
35
HBM
A2540-1.1 en/de
11
Parameter factory settings
When the DIS2116 is delivered, it comes with specific factory settings:
•
The (7-digit) production number is entered
As the DIS2116 has undergone certain tests during production, the following
information can deviate from zero:
•
Legal-for-trade counter (Menu ”INFORMATION SCALE” or F1 key with
symbol “i”)
•
Sensor-overflow counter (Menu ”INFORMATION BUSSCAN”)
Parameter
Default value
Explanation
Limit values 1...2
Deactivated
Filter mode
0
Standard filter
Filter
5
0.3 Hz (AD104C, C16i)
Manual tare value
0
Deactivated
COM1, baud rate
38400
COM1, parity bit
even
COM2,COM3,COM4, Function
OFF
Deactivated
COM2,COM3,COM4,baud rate
9600
COM2,COM3,COM4,parity bit
even
Printing, ESC sequences
0
Deactivated
Basic scale function
Standard
Non-automatic scale
Menu access
5
All menus
Legal-for-trade operation
OFF
Deactivated
Unit
OFF
No unit of measurement
Zero on start-up
OFF
Deactivated
Zero tracking
OFF
Deactivated
Standstill monitoring
OFF
Deactivated
Increment
1d
Decimal point
0
right
Maximum capacity
10000
Multi-range scale
0
Deactivated
Calibration weight
10000
Max. capacity adjustment
Gravitational acceleration
correction
9,8102
Deactivated
Scale char. curve, zero
0
Scale char. curve, nom. value
10000
At load cell maximum capacity
Linearization
Deactivated
When restoring the factory settings (”FACTORY SETTINGS DIS2116” menu),
the parameters described above are set to their default values.