8.12.1 Tool data
Function
A maximum of 10 tool data sets with defined tool-specific
parameters can be stored in the unit.
Tool-specific parameters
A tool data set comprises the following parameters:
Parameter
Comment
Tool No.
Tool name, max. 7 characters
Nominal value 1
Upper dev. nominal/actual
Lower dev. nominal/actual
Difference main/return line
Difference main
line/external
Flow rate internal max.
Flow rate internal min.
Save tool data
Fig. 68: Save tool data
Proceed as follows in order to save the current limit value settings
(temperature and flow rate) as well as nominal value 1 to the
selected tool data set:
1.
Display menu page
Monitoring \ Tool data \ Tool 1..10
.
2.
Select the parameter
Save tool data
and confirm with the
key.
The limit values (temperature and flow rate) and nominal
value 1 are saved to the selected tool data set.
Load tool data
Fig. 69: Load tool data
Proceed as follows in order to load the selected tool data set as
limit values (temperature and flow rate) or nominal value 1:
1.
Display menu page
Monitoring \ Tool data \ Tool 1..10
.
2.
Select parameter
Load tool data
.
3.
Press the
key.
The parameters from the selected tool data set are loaded as
limit values or nominal value 1.
Parameters that are set to "--" will not be loaded.
NOTE!
When loading tool data, monitoring is automatically
set to "manual".