HB-100/140/160Z1
Temperature Control Unit
Operation
94
O8285-EN 2017-12
Loading data
Fig. 67 Loading data
Proceed as follows in order to load data to the unit from a USB
data carrier:
1.
Display menu page
Save/Load
.
2.
Connect USB data carrier to front connector.
3.
Select the data to be loaded and confirm with the
key.
4.
In the Explorer window, select the directory and file and
confirm with
.
The data is loaded to the unit. If loaded values are outside the
permissible range, then these are reset to the standard
settings.
Serviceinfo
Exa.
Serviceinfo_2017-03-10_15-26-08
Time
Date
Configuration data
Exa.
HB 160 Z1 8 FU A2 400 1 [1].csv
Index
1
Address
Mains voltage
Cooling type
Pump type
Heating type
System and housing size
max. operating
temperature
Parameter data
Exa.
Par HB 160 Z1 1 [1].csv
Index
1
Address
System and housing size
max. operating temperature
Error and Operation data
Exa.
BD HB 160 Z1 1 [1].csv
Index
1
Address
System and housing size
max. operating temperature
1
An index is automatically added when the file name already exists.
File name
The unit automatically creates file names on the USB data carrier
according to the following examples: