Sequestering on the Cell Saver 5+ System
P/N 53063-50, Manual Revision: B
7-18
On the first cycle the bowl size is determined by the machine. The high volume
bowl (225 ml) and low volume bowl (125 ml) can both be used for sequestering.
By default the machine expects to detect a low volume bowl (125 ml). If on the
first cycle, the machine detects a high volume bowl just prior to collecting PPP,
the pump is paused and the display prompts the operator to confirm that a high
volume bowl is being used. The following display is shown:
The machine expects a response from the operator within 15 seconds. If a low
volume bowl is being used, the operator can override the machine detector and
select a low volume bowl by pressing the NO key. If a high volume bowl is being
used the operator can press the YES key or nothing. Once a key is pressed or 15
seconds has elapsed the pump restarts and the machine continues processing.
7. Plasma is being collected in the PPP bag. The display is incrementing the
PPP volume along with the Processed Volume and displays PPP Collection
phase as shown below:
PAUSE
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press YES to select HIGH volume bowl
SEQUESTER
215 ml
0 ml
0
Pump Rate:
0
Centrifuge:
4750
Press NO to select LOW volume bowl
ml/min
rpm
PPP Volume:
PRP Volume:
0 ml
0 ml
0/
0/
HIGH Volume Bowl sensed
*Is this correct?
Y/N?
125
ML
PPP
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press SELECT to advance highlight
SEQUESTER
166 ml
0 ml
0
Pump Rate:
60
Centrifuge:
4750
Press ARROW keys to change values
ml/min
rpm
PPP Volume:
PRP Volume:
27 ml
0 ml
27/
0/
Press MODIFY to change parameters