P/N 53063-50, Manual Revision: B
Sequestering on the Cell Saver 5+ System
7-9
If the “Return Plasma?” parameter in the Sequester Setup is set to “YES” the
following screen is displayed:
7. When the START key is pressed, the pump will speed up to 400 ml/min to
pump the RBCs from the bowl to the reinfusion bag. Under certain
conditions the centrifuge will briefly rotate in both directions to expedite the
removal of the red cells from the bowl. The following display will be shown:
125
ML
EMPTY BOWL
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press START to empty bowl
SEQUESTER
910 ml
236 ml
2
Pump Rate:
0 ml/min
RBC
Press MODIFY to change parameters
OPEN PPP bag clamp
CLOSE PRP bag clamp
125
ML
EMPTY BOWL
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press ARROW keys to change values
SEQUESTER
910 ml
236 ml
2
Pump Rate:
0 ml/min
RBC
Press MODIFY to change parameters
Содержание Cell Saver 5+
Страница 2: ...Preface P N 53063 50 Manual Revision B ii...
Страница 134: ......
Страница 154: ......