P/N 53063-50, Manual Revision: B
Sequestering on the Cell Saver 5+ System
7-15
Milk Delay:
The time in seconds the pump will be stopped during the PRP
collection phase before starting to pump the programmed Milk Vol parameter.
2. Once the Assisted Sequester Setup parameters have been set and the START
key is pressed to begin a new procedure, the screen display changes to the
STANDBY screen and further information is requested before processing
can begin. The machine needs to know which bag to manage the system air
in. The display is as follows:
Selecting the YES key uses the PPP collection bag to manage the air, while
pressing the NO key selects the air bag.
3. Once a selection is made for system air management, the display changes to
the following to assist the operator in final setup confirmation of the
disposable set.
If the PPP collection bag was selected to manage the system air:
STANDBY
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press YES to use the PPP collection bag for air
SEQUESTER
Press NO to use the air bag for air
PPP Volume:
PRP Volume:
Use PPP bag for Air?
0
0
0
0
0
ml
ml
ml
ml
Y/N
0/
0/
STANDBY
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press START to fill bowl or CONC to concentrate
SEQUESTER
Press MODIFY to change parameters
PPP Volume:
PRP Volume:
0
0
0
0
0
ml
ml
ml
ml
0/
0/
Press YES to select Cell Saver protocol
CLOSE RED bag clamps
CLOSE AIR bag clamp
OPEN YELLOW bag clamp
CLOSE BLUE bag clamp
**AGITATE BLOOD BAG**