P/N 53063-50, Manual Revision: B
Maintenance
8-5
Filter Cleaning Procedure
Cleaning the air filter is simple and easy. The following steps outline the correct
procedure.
1. Unplug the CS5+.
2. If there is a louvered panel covering the filter, remove the four screws which
hold it in place, and set the panel and screws aside.
3. Grasp the filter and pull it out.
4. Rinse the filter under warm running water until it is clean. Do not use soap
or any cleaning solution.
5. Squeeze out excess water, place on a clean cloth, and allow to dry
completely.
6. Ensure the filter is 100% dry. Reinsert the filter into the panel, making sure
the filter completely covers the opening. Replace the louvered panel, if so
equipped, with the openings facing down.
7. Record the date of maintenance.
Note: A quick reference guide for cleaning the Cell Saver 5+ is found in
Appendix B.
Содержание Cell Saver 5+
Страница 2: ...Preface P N 53063 50 Manual Revision B ii...
Страница 134: ......
Страница 154: ......