P/N 53063-50, Manual Revision: B
Setup Procedures for Cell Salvage
4-11
Warning: The bowl base (or adapter) must be firmly installed and evenly seated
in the centrifuge chuck. If the centrifuge chuck spins with the bowl base (or
adapter) not evenly seated, as indicated by bowl wobbling or noise, bowl
damage will occur and the bowl and harness set must be replaced.
Once the disposable is properly loaded, press START. A screen similar to the
following will appear.
Note: Press YES to restore default settings.
This screen notifies the user of the current settings of the modifiable parameters.
These parameters may be changed by pressing the MODIFY key.
Similar screens appear when using the 70 ml bowl.
CURRENT SETTINGS
Press START to begin procedure
Press MODIFY to change settings
Min Wash Vol. - 225:
-125:
Reservoir Level:
Resume at Level:
Reservoir Level Sensor:
Pump Regulation:
Empty RBCs to:
AUTOMATIC
Press YES to restore default settings
NO WASH
NO WASH Option:
Protocol:
OFF
Cell Saver
1000 ml
750 ml
800 ml
400 ml
ON
ON
Blue Line
CELL SAVER SETUP
Press ARROW keys to change values
NO WASH Option:
Protocol:
OFF
Cell Saver
Min Wash Vol. - 225:
-125:
Resume at Level:
Reservoir Level Sensor:
Pump Regulation:
Empty RBCs to:
AUTOMATIC
750 ml
400 ml
ON
Blue Line
Reservoir Level:
800 ml
ON
1000 ml
Press SELECT to advance highlight
Press MODIFY when finished