P/N 53063-50, Manual Revision: B
Automatic Operation: Cell Salvage
5-19
S
YSTEM
STATUS
The next section briefly discusses some information regarding the display and
how current procedure statistics are displayed.
The current system status is displayed as shown in the following example:
The top left section displays the current operating status of the machine. In this
case, the CS5+ is in the FILL mode and fluid is being pumped from the reservoir
into the bowl.
The right shows the current procedure statistics and parameters. In this case, the
CS5+ is under AUTO control, the pump is filling at 600 ml/min, 2 bowls have
been processed, 2000 ml of saline solution have been used, a total of 2525 ml of
volume has been processed from the reservoir and 450 ml of concentrated RBCs
have been sent to the reinfusion bag.
To increase the pump speed
temporarily,
press the UP (+) slew key. This allows
the operator to make minor adjustments to speed without reprogramming the
system. These temporary changes are valid only for the present mode (e.g., FILL)
and as the system enters the next mode, the programmed values take over again.
The temporary pump speeds may be increased to 600 ml/min (300 ml/min for the
70 ml bowl) or decreased to 200 ml/min.
Warning: The effluent line sensor is disabled with any pump speed adjustment
by the operator.
225
ML
FILL BOWL
Processed Vol:
Wash Vol:
Reinfusion Vol:
Bowls Processed:
Press MODE twice to enter EMERGENCY option
AUTOMATIC
2525 ml
2000 ml
450 ml
2
Pump Rate:
600 ml/min
Содержание Cell Saver 5+
Страница 2: ...Preface P N 53063 50 Manual Revision B ii...
Страница 134: ......
Страница 154: ......