P/N 53063-50, Manual Revision: B
Sequestering on the Cell Saver 5+ System
7-13
M
ETHOD
2 — A
SSISTED
SEQUESTER
MODE
1. Collect whole blood from the patient via short intravenous or arterial
cannula into anticoagulated blood bag(s).
Note: Only one unit of blood should be collected and processed at a time. Unless
directed by a physician, do not begin withdrawal of the second unit of blood from
the patient until the first unit of concentrated red autologous blood cells has been
returned.
Power on the machine and select the Sequester protocol at the Cell Saver SETUP
screen. If the Assisted Sequester Mode parameter is set to OFF, use the SELECT
key to highlight it and the ARROW keys to select Assisted Sequester Mode to YES.
The following SEQUESTER SETUP screen is displayed:
The following table indicates the modifiable parameters in the Sequester mode of
those machines equipped with Assisted Sequester. The table indicates the default
setting of each parameter, the maximum and minimum values attainable, and the
incremental value when it is changed by pressing the up or down arrow key.
Note: The “milk delay” parameters will only be displayed if the “milk PRP
collection layer?” parameter is set to YES. “Milking” Enables an automatic pump
start/stop/restart sequence during collection.
SEQUESTER SETUP
Press ARROW keys to change values
Press SELECT to advance highlight
Assisted Sequester Mode:
Fill Bowl DRAW Rate:
Fill Bowl Centrifuge Speed:
PRP Collection DRAW Rate:
PRP Collect Centrifuge Speed:
PRP Extend Collection Volume:
ASSISTED
Press MODIFY when finished
60
30
Return Plasma?
Milk PRP Collection Layer?
Milk Vol: 3 ml Milk Delay:
4750
NO
YES
10
ON
60
4750
ml/min
rpm
ml/min
rpm
ml
sec