P/N 53063-50, Manual Revision: B
Equipment and Disposable Description
2-13
Disposables
A variety of disposables are available for use with the CS5+. Chapter 4 explains
the available disposables in detail. The purpose of this section is to give a general
understanding of the uses and operating characteristics of the disposables rather
than to provide specifications:
Figure 2-8: Cell Saver 5+ disposable harness with Latham bowl
Harness
Figure 2-8 illustrates the harness used with the CS5+. The harness consists of
three lines:
1. the yellow line from the wash solution,
2. the red line from the unprocessed blood source,
3. the blue line to the reinfusion bag.
All three lines pass through valves on the CS5+. The three lines are joined at a
tubing manifold (see Figure 2-8). A clamped line sensor is located under the blue
line to warn of occluded lines during EMPTY and RETURN. After the junction,
the single line passes through the pump and then the air detector. After passing
through the air detector, the single line enters the centrifuge well through a slot
in the rim where it is held down by the cover. Once inside the well, the line is
attached to the inlet port of the bowl.
The effluent line is attached to the outlet port of the bowl and exits the centrifuge
through the effluent line sensor and is then connected to the waste bag.
10
9
8
7
6
5
4
3
2
1
HAEMONETICS
DO NOT USE
WITH PRESSURE CUFF
MILLILITERS
(APPROXIMATE)
IOOO
9OO
8OO
7OO
6OO
5OO
4OO
3OO
2OO
IOO
Reinfusion Bag
Effluent Line
Waste Bag
Blue
Line
Saline Bag
Spikes
Collection
Reservoir
Connector
Yellow Line
Red Line
Tubing
Manifold
Bowl