Automatic Operation: Cell Salvage
P/N 53063-50, Manual Revision: B
5-20
To return to full automatic operation (enabling the effluent line sensor):
1. Press the UP (
▲
) or DOWN (
▼
) key until the “Pump Regulation Off”
message disappears from the screen. This will occur when the pump is
returned to its original speed.
OR
2. Press the PUMP/PAUSE key twice, once to pause the pump and a second
time to resume at the default speed.
Wash Volume
Monitor
During any automatic WASH mode, the display panel displays a running,
cumulative total of wash solution used on the current WASH mode on the left
hand side of the display, as shown in the following figure. This counting function
returns to zero each time the WASH mode is entered, and terminates when the
machine advances to EMPTY. An indication of the Total Wash volume used is
shown in the procedure statistics area on the right side of the display as indicated
in the following figure.
225
ML
WASH CELLS
Processed Vol:
Wash Vol:
Reinfusion Vol:
Bowls Processed:
Wash Volume: 150/1000 ml
Press MODE twice to enter EMERGENCY option
AUTOMATIC
800 ml
150 ml
0 ml
0
Pump Rate:
500 ml/min
NaCl
Press MODIFY to change volume of wash