Sequestering on the Cell Saver 5+ System
P/N 53063-50, Manual Revision: B
7-22
12. Once the pump has pumped the programmed PRP Extended Collection
Volume, the pump is stopped and the PRP collection is complete and the
display adds COMPLETE to the screen as follows:
13. Press the EMPTY key to stop the centrifuge. When the centrifuge comes to a
complete stop, the screen display will prompt the operator to close the blue
(PRP) collection bag clamp and open the bag clamp that has the system air.
Once this is completed the START key must be pressed to empty the
contents of the bowl.
If the “Return Plasma?” parameter in the Sequester Setup is set to “NO” the
following screen is displayed:
125
ML
PRP
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press START to continue collection into PRP layer
SEQUESTER
299 ml
0 ml
0
Pump Rate:
0
Centrifuge:
4750
Press EMPTY to end collection of PRP
ml/min
rpm
PPP Volume:
PRP Volume:
100 ml
60 ml
100/
60/
Press MODIFY to change parameters
COLLECTION
COMPLETE
125
ML
EMPTY BOWL
Processed Vol:
Reinfusion Vol:
Bowls Processed:
Press START to empty bowl
SEQUESTER
299 ml
0 ml
0
Pump Rate:
0 ml/min
RBC
Press MODIFY to change parameters
OPEN AIR bag clamp
CLOSE BLUE bag clamp
PPP Volume:
PRP Volume:
100 ml
60 ml
100/
60/