58
Standard Operations
5.5.6.3 Program a new test
1.
The instrument needs to be at the Main Menu.
2.
Insert the sample cuvette/vial or blank (depending on the
working procedure) into the cell compartment #1 and close the
cell compartment.
3.
After "Program not available", a data overview is displayed,
including the specifications of the test that is to be revised.
The working procedures of the corresponding test contains the
new data (Wavelength, Factors, Measurement Ranges,
Conversion Factor for chemical Form etc.).
4.
Highlight the corresponding line including the specifications of
the test that is to be programmed and press
Edit
.
5.
To finish programming, press
Store
and then
Cancel
to return
to the Main Menu.
6.
Press
Start
to run the test with the new data.
5.5.7 Upgrade of the instrument software
To obtain the software for the update from the Internet at
www.hach.com
:
1.
Go to
http://www.hach.com
.
2.
On the DR 5000 product page, click
Lab System
Software
/
Software Update Downloads
under the heading
Downloads
.
3.
Locate the appropriate download and follow the prompts for
saving the file(s) to the USB memory stick.
4.
On the DR 5000, press
Instrument Update
in the System
Checks menu.
5.
Connect the USB stick to the USB interface on the DR 5000
(
section 3.3 on page 14
). Press
OK
.The link is established
automatically and the software is updated.
Содержание DR 5000
Страница 2: ...2 ...
Страница 6: ...6 Table of Contents ...
Страница 8: ...8 Specifications ...
Страница 12: ...12 General Information ...
Страница 20: ...20 Start Up ...
Страница 60: ...60 Standard Operations ...
Страница 120: ...120 Advanced Operations ...
Страница 128: ...128 Troubleshooting ...
Страница 130: ...130 Replacement Parts ...
Страница 132: ...132 Contact Information ...
Страница 134: ...134 Limited Warranty ...