93
Advanced Operations
6.5.2 Complete a measurement in the multi wavelength mode
1.
Insert the blank cuvette/cell into the cuvette/sample cell holder.
Press
Zero
.
Note:
The
Read
key does not become active until the zero
measurement has been completed.
2.
Insert the sample cuvette/cell into the cuvette/sample cell
holder. Press
Read
.
Note:
The UV lamp is not switched on until a wavelength in the UV
spectrum has been entered and the blank reading has been started.
During the warm-up phase of the UV lamp, the message ”Warming
up...“ is displayed and the ”UV lamp“ symbol flashes. As soon as the
UV lamp is ready, the blank reading is carried out.
Note: Zero
and
Read
are disabled until the cell compartment is
closed.
3.
For data storage, see
section 5.3.1.1 on page 35
.