45
Standard Operations
5.4.4.1 Take single wavelength measurements (single reading)
1.
Insert the blank cuvette into the cuvette/sample cell holder.
Press
Zero
.
Note:
The
Read
key is only active after the zero measurement has
been carried out.
2.
Insert the sample cuvette into the cuvette/sample cell holder.
Press
Read
.
Note:
For data storage, see
section 5.3.2 on page 38
.
5.4.4.2 Take single wavelength measurements (continuous readings)
1.
Insert the blank cuvette into the cuvette/sample cell holder.
Press
Zero
.
Note:
In the reading mode ”Continuous“ only the Zero key is shown to
start the reading. The reading sequence is started automatically.
2.
Insert the sample cuvette into the cuvette/sample cell holder.
3.
Press
Options
and then the
Store
icon to store the displayed
data in the Data Log.
Note:
For data storage, see
section 5.3.2 on page 38
.
5.4.5 Set the dilution factor
The Dilution Factor function is used to calculate the original
concentration of a sample that has been diluted by a known ratio.
For example, dilution factor can bring the analyzed concentration
within the test range.
1.
Press
Options
>
More…>Dilution Factor
.
The number entered at the dilution factor prompt will be
multiplied by the result to compensate for the adjustment.
For example, if the sample has been diluted by a factor of 2,
enter 2. The default setting of the dilution factor is turned off.
2.
Press
OK
to confirm.
Содержание DR 5000
Страница 2: ...2 ...
Страница 6: ...6 Table of Contents ...
Страница 8: ...8 Specifications ...
Страница 12: ...12 General Information ...
Страница 20: ...20 Start Up ...
Страница 60: ...60 Standard Operations ...
Страница 120: ...120 Advanced Operations ...
Страница 128: ...128 Troubleshooting ...
Страница 130: ...130 Replacement Parts ...
Страница 132: ...132 Contact Information ...
Страница 134: ...134 Limited Warranty ...