102
Advanced Operations
Time & interval:
1.
Press
Time & Interval
in the Options menu.
2.
Input the total time and the reading time and press
OK
to
confirm.
Note:
In total 500 measuring steps are possible. To select a total time and
a time interval that would cause this number of measurements to be
exceeded, the time interval is defined automatically and the
OK
key is
inactivated.
Scale & units:
1.
Press
Scale & Units
in the Options menu.
2.
Highlight
Abs
or
%T
as the required units.
3.
Highlight
Auto
or
Manual
scaling on the graph´s y-axis.
Note:
If manual scaling is selected, use the alphanumeric keypad to set the
limits y
min.
and y
max.
. The graph is adjusted to display only the values in the
selected range. If automatic scaling is selected, the instrument sets the
limits automatically so that the total range can be displayed.
4.
Press
OK
to confirm.
5.
Press
Return
to return to the scan mode.
6.7.2 Time course scan reading
After the parameters have been selected, the instrument must be
blanked, then the sample can be analyzed.
1.
Insert the blank cuvette/cell into the cuvette/sample cell holder.
Press
Zero
. The blank reading is shown on the display.
2.
Insert the sample cuvette/cell into the cuvette/sample cell
holder. Press
Read
. Start collecting time course (kinetic) data.
Note:
During the measurement the
Zero
and
Read
keys change to
Mark
and
Stop
.
Содержание DR 5000
Страница 2: ...2 ...
Страница 6: ...6 Table of Contents ...
Страница 8: ...8 Specifications ...
Страница 12: ...12 General Information ...
Страница 20: ...20 Start Up ...
Страница 60: ...60 Standard Operations ...
Страница 120: ...120 Advanced Operations ...
Страница 128: ...128 Troubleshooting ...
Страница 130: ...130 Replacement Parts ...
Страница 132: ...132 Contact Information ...
Страница 134: ...134 Limited Warranty ...