115
Advanced Operations
8.
Insert the different filter in the given order one after the other
with the filter glass facing the front of the instrument. Press
Next
after inserting a filter and closing the cell compartment
After the last measurement the results are displayed.
9.
Press
PC & Printer
icon to send the data to a USB memory
stick, PC or to a printer.
The files will be stored automatically as CSV file (Comma
Separated Value). The file name will be formatted as
“Verification.csv“.
6.8.4 Output checks
If a printer is connected a test printing of the current screen will be
printed.
6.8.5 Lamp history
The Lamp History menu provides the following information:
•
The amount of time that the lamp has been on (Hours)
•
The number of times that the lamp has been turned on (Cycles)
•
The number of times that the instrument tried to turn the UV
lamp on (triggers)
After a lamp is replaced and the Lamp History is reset, the display
of the total operating time is reset to 0.
1.
Press
Lamp History
in the System Checks menu.
2.
Press
Reset VIS
and the Visible Lamp will be reset.
3.
Press
Reset UV
and the UV Lamp will be reset.
4.
Press
OK
to return to System Checks.
Содержание DR 5000
Страница 2: ...2 ...
Страница 6: ...6 Table of Contents ...
Страница 8: ...8 Specifications ...
Страница 12: ...12 General Information ...
Страница 20: ...20 Start Up ...
Страница 60: ...60 Standard Operations ...
Страница 120: ...120 Advanced Operations ...
Страница 128: ...128 Troubleshooting ...
Страница 130: ...130 Replacement Parts ...
Страница 132: ...132 Contact Information ...
Страница 134: ...134 Limited Warranty ...