112
Advanced Operations
7.
The result is displayed and is to be compared with the
nominal/standard data of the test filter. If a specific sample is
used, the measured results must be evaluated against
user-defined requirements.
8.
Press
Cancel
to return to Optical Checks.
6.8.3.5 Drift check
The Drift Check test is used to test the stability of the instrument.
Note:
The Drift Check runs 1 hour.
1.
Press
Drift Check
in the Optical Checks menu.
2.
Press
λ
to input the wavelength.
3.
Input the wavelength and press
OK
to confirm.
4.
Press
Start
.