5.3.2
Switch off procedure model range 200/75 up to 450/300
no. Illustration
Description
[01]
To switch into “stand by” mode press the black “OFF”
push button.
[02]
Then switch off the running machine turn the red main
switch at the right side of the external control unit.
The main switch has to be turned 90° counter
clockwise. The main switch can be secured against
unauthorized access with a padlock.
N
The ON-/OFF intervals must not be shorter than 1-2 minutes (for models with 3-phase
connection: 2-3 minutes). This may otherwise lead to interference and/or failure of the backup
and/or data loss and/or data corruption.
N
Models for medical application with the isolation transformer have an inrush current limiter
(surge guard). Too short ON/OFF intervals will lead to a deactivation of the inrush current limiter
and result in an overload of the circuit fuse.
N
If the device is used daily, we recommend switching it on in the morning and leave it in stand-by
mode during the day.
N
Press the emergency stop only in case of danger.
N
The emergency stop is not to be used as a normal stop key.