operation
file: n:\article\cos14310m5-v1_07hpc-en\20130207_cos14310m5-v1.07hpc-en_h-p-cosmos_treadmill_manual.doc
© 2013 h/p/cosmos sports & medical gmbh author: fh [email protected] created 07.02.2013 printed 04.03.2013 page: 50 of 197
If the running machine does not receive a heart rate signal when starting the cardio mode, it will start with the set
starting speed and will not alter this speed. Every 30 seconds a warning signal is audible, after 1 minute the
running machine cancels the mode (at MCU5 firmware-versions < V1.03.1 or MCU4 firmware-versions < V4.04.2,
the running machine cancels after 2 minutes). If the heart rate signal fails during the operation in the cardio mode,
the running machine will continue running in the actual set speed and angle of inclination. In this case, beeper alert
signal will be activated every 30 seconds. The running machine will run down speed to 2 km/h and the mode will
be terminated after 1 minute (at MCU5 firmware-versions < V1.03.1 or MCU4 firmware-versions < V4.04.2, the
running machine chancels after 2 minutes).
Examples:
a) Set max. speed 4.0 km/h if you want the subject only to walk and to control the load by means of elevation.
b) Set max. speed 20.0 km/h if you want the subject to control the load by means of speed only without elevation.
The running machine adjusts speed and elevation according to the following matrix.
Chart for heart rate-controlled work load at the heart rate lower level
Heart rate difference actual > < set value
Speed
(km/h)
Elevation
(%)
Time
(s)
0 ... 5
0.2
0.1
25
5.1 ... 15
0.4
0.2
25
15.1 ... 30
0.6
0.4
25
30.1 ... 50
0.8
0.8
20
> 50
1.0
1.0
20
Chart for heart rate controlled work load at the heart rate upper level
Heart rate difference actual > < set value
Speed
(km/h
Elevation
(%)
Time
(s)
0 ... 5
0.3
0.3
12
5.1 ... 15
0.8
0.8
12
15.1 ... 30
1.0
1.0
10
30.1 ... 50
1.5
1.2
8
> 50
2.0
1.6
7
The following table show the operating steps to run the treadmill via firmware or software h/p/cosmos para control
®
Operation of cardio mode
Initiation: Running-Belt is not moving. One of the modes LED
is flashing: (manual, profile, cardio, test)
Step no. Activity
Keys
Response / Display
[01]
Select cardio mode
a
or
d
Alteration of the mode (manual, profile, cardio, test)
until
cardio is flashing
[02]
Start cardio mode
c
cardio is illuminated,
speed max. is illuminated
=
indicates 6.0 is (for max. speed during cardio mode).
This default value can be changed in the optional settings.
% is illuminated,
energy is illuminated
[03]
Change max. speed
for cardio mode
a
or
d
=
indicates flashing: 2.0 ... max. speed of the running
machine (for max. allowed speed during cardio mode)
Examples:
a) Set max. speed 4.0 if you want the subject only to walk and
to control the load by means of elevation.
b) Set max. Speed 20.0 if you want the subject to control the
load by means of speed only without elevation.