optional settings: administrator options
file: n:\article\cos14310m5-v1_07hpc-en\20130207_cos14310m5-v1.07hpc-en_h-p-cosmos_treadmill_manual.doc
© 2013 h/p/cosmos sports & medical gmbh author: fh [email protected] created 07.02.2013 printed 04.03.2013 page: 139 of 197
OP38 Service
interval
(distance-dependent)
"E 02" "HELP"
h/p/cosmos recommends regular maintenance (interior cleaning,
checking drive belts, etc.) carried out by an authorized technician at
least every 5,000 km (for ladder ergometers: 500 km). For this
reason, the error message "E 02" "HELP" is displayed every 5000 km
(500 km). For detailed information, descriptions and specifications
about maintenance, please contact the h/p/cosmos service
department. An error message "E 02" "HELP" for the maintenance
interval will appear. TSC controls are to be performed at least every
12 months.
Reset of the interval and the error report „E02“ is possible via option
OP47.
The service maintenance interval can be changed via option OP38.
Call up OP38 and change with
a
or
d
.
Confirm with
c
.
§
reports: service interval in km (standard = 5.000 / 500)
Adjustment range 1,000 to 9,900 km or 100 to 5,000 km for the ladder
ergometer.
OP39
Elevation: display as set point or
true value
(This option is only available for
treadmills, not for ladder
ergometers)
0 = Display of elevation as set point
1 = Display of elevation as true value resp. flashing set point report, if
the set point was changed with the keys and has not yet been
reached (report: current elevation angle; recommended default
setting).
OP40
Speed: display as set point or true
value
0 = Display of speed as set point (report: target speed)
1 = Display of speed as true value resp. flashing set point report if
the set point was changed with the keys and has not yet been
reached. (report: current speed; recommended default setting)
OP41
Elevation control after STOP
(This option is only available for
treadmills, not for ladder
ergometers).
0 = Elevation remains at the present level after pressing the STOP
button (recommended; default-setting)
1 = Elevation decreases to 0 % after pressing the STOP button.
Setting of OP 41 to value 1 affects the safety of the subject and third parties.
A downward movement of the elevation after pressing STOP can result in a jumpy
reaction and in severe injuries and contusions.
h/p/cosmos strongly recommends adhering to the standard setting of this option. Stop
means stop for all movements and must not result in activity or movement on the part of
the running machine. It is only allowed to change option 41 after a written documented
instruction of the user regarding the danger warnings and after a written confirmation by
the customer/user.
OP46
Display and setting of elevation for
ladder ergometer
(only available for ladder
ergometers, not for treadmills)
$
indicates elevation of ladder ergometer, standard: 75 degree,
adjustable from 0 to 180 degrees
OP47
Reset of oil and service interval
After service including lubricating of the treadmill and all maintenance
work, the oil and service intervals can be reset with option OP47.