optional settings: administrator options
file: n:\article\cos14310m5-v1_07hpc-en\20130207_cos14310m5-v1.07hpc-en_h-p-cosmos_treadmill_manual.doc
© 2013 h/p/cosmos sports & medical gmbh author: fh [email protected] created 07.02.2013 printed 04.03.2013 page: 143 of 197
Standard settings of administrator options
Option Description
Standard setting
Adjustable range
OP 01 Reset of error messages E20, E21,
E30, E31, E32, E41, E50 and E51
OP 02 OEM code
depending on the model
0 = h/p/cosmos (Standard)
1 = JAEGER / VIASYS / Cardinal
Health / CareFusion
2 = Ergo-Fit (1997-98)
3 = Proxomed / Kardiomed
4 = KISTLER Gaitway
6 = COSMED
9 = SCHILLER
OP 03 Selection of device type:
see list of running machines types
depending on the model
depending on the model
OP 04 Maximum speed for reverse belt
rotation
depending on the device type
in OP03
1.0 km/h .... max. speed
OP 05 Maximum speed
depending on the device type
in OP03
1.0 km/h .... max. speed
OP06
Max. acceleration time in seconds
Increment 131 = 5 sec.
Increments: 255 ... 1
Seconds: 2.6 ... 655.4
OP 07 Maximum elevation
depending on the device type
in OP03
0 .... max. elevation
OP 08 Heart rate display interval
0 = beat to beat
0 or 1 ... 9 seconds
OP 09 Display test for all displays
OP 10 Enter the serial number of the
device
300…
12-digit serial number
OP 11 Activating and deactivating speed
measurement
depending on the device type
in OP03
0 = without speed measurement
1 = with speed measurement
OP 12 Setting of allowance for speed signal
comparison
(measurement / control)
6 %
This value does not reflect the
speed accuracy of the running
belt.
0 = no comparison
6 = 6 % deviation allowed
8 = 8 % deviation allowed
10 = 10 % deviation allowed
OP13
Motor brake
OFF
OFF = not active
ON = active
OP 14 Maximum acceleration time for
RS232
Increment 41 = 16 sec
Increments: 255 ... 1
Seconds: 2.6 ... 655.4
OP 16 Activating and deactivating heart
rate measurement
(depending on the model)
0 =
HF measurement deactivated
1 =
HF measurement activated,
HFU1 + RE06REC
2 =
HF measurement activated,
HFUi;
auto = autom. detection
OP 33 Distance between rungs (only for
ladder ergometer)
254 = 25.4 cm
Distance between rungs in 0.1 cm
OP 34 Manual speed calibration
(depending on the device
type, OP 03)
Range from: 0 to 65535
OP 35 Oil interval
1.000 km (treadmills)
100 km (ladder ergometer)
0 = no supervision
100 ... 5.000 km