20
4.4 Testing the device by tester. (Two weeks test again)
User can use the tester to check if the device work properly, specific
operation is as follows:
a. Pulling out the measure wire and turn off the device.
b. Insert the tester
c. Turn on the device and press the En-Ap button. Then you can enter
into apex measure module. The screen shows “01” “02”or ”03” or”04”.
It means the machine is fine.
4.5 Operatio
n
instruction
4.5.1 Please let the measure wire insert the base. Then staring up
the power button. Next press the “En-Ap” button. And you can enter
into the apex mode.
4.5.2 When the device is starting up, you can press the power button
again. And the machine can turn off.
4.5.3 If you press the“ VOL” button, you can turn up the voice.
4.5.4 Hang the lip hook on the lip, make sure it contact the oral
mucosa as a reference electrode [Picture 5].
4.5.5 Clip the file with file clip, approach to the apex, then there will
be continuous alarm when the distance is less than 2mm [Picture 6].
Picture 5 Picture 6