25
Problem
Cause
Solution
After the motor
handpiece is
activated, if
the wireless
communication
connection status
indication is
“
”, the
wireless
connection
between the
motor handpiece
and the base
fails.
1. The wireless
connection has failed
2. Handpiece far from
the base.
1.Press the handpiece “ON/
OFF” button more than 5
seconds to power off and
power on again.
2. Place the handpiece near
the base and power on.
The contra-
angle cannot be
calibrated.
The calibration
procedure may have
been interrupted by
increased resistance in
the contra-angle.
If calibration has been
interrupted, calibrate the
motor handpiece again to rule
out the possibility of a motor
fault.
2.Clean and lubricate the
contra-angle.
3.Start the calibration
procedure again.
Motor handpiece
is getting hot.
Run time is too long
with reciprocating
mode
Allow the device to cool
down and start again the
motor handpiece.
Continuous
rotary file blocks
in the root canal
Wrong file setting.
Too much pressure on
the instrument.
Change the rotational
direction by pressing the
REV key. Start the handpiece
and pull out the file carefully.
Reciprocating
file blocks in the
root canal.
Too much pressure on
the instrument.
File not frequently
clened.
Try to remove the file with a
pair of pliers by pulling out
and rotating the file gently
clockwise.