36
the running properties can change with usage, cleaning and sterilization.
See section 3.8 for details.
8.3.6 Battery replacement
8.3.6.1 When the battery of the base or motor handpiece cannot be
charged, or the base or the motor handpiece cannot be switched on and
the battery needs to be replaced, it must be sent back to the manufacturer
or replaced by the service personnel authorized by the manufacturer.
8.3.6.2 Battery replacement method:
a) Turn the base or the motor handpiece power off.
b) Use tweezers etc. to open the rubber cover and then remove the
screw.
c) Remove the battery cover.
d) Remove the old battery and disconnect the connector.
e) Connect the new battery and put it in the base or the motor
handpiece.
f) Replace the cover and its screw.
It is recommended to contact local distributors or manufacturer to
replace the battery.
9 Environmental protection
Please dispose according to the local laws.
10 After service
From the date this equipment has been sold, based on the warranty
card, we will repair this equipment free of charge if there are quality
problems. Please refer to the warranty card for the warranty period.
11 Symbol instruction
Date of manufacture
serial number
Type B applied part
Manufacturer
CE marked product
Class
Ⅱ
equipment
Follow Instructions
for Use