19
b. Make sure if the plug of the measuring wire is inserted into the
socket correctly.
c) Make sure if the file clip and lip hook are connected well to the
measuring wire.
d) Make the lip hook touch the bent wire of the file clip [as showed
in picture 3] to confirm all the instruction bars are displayed on the LCD
screen and static display the digital ‘-3’, otherwise , it means that the file
clip or the measuring wire is damaged, should be replaced.
Picture 3
Picture 4(
a
) Picture 4(
b
) Picture 4(
c
)
4.3 Explanations on the interfaces displayed
a) The screen displays the front region of the apical foramen by
instruction bars. Please refer to the green region as showed. [Picture 4(a)]
b) The file has gone to the position near by the apical foramen when
it comes to the orange bars. [Picture 4(b)]
c) The file has exceeded the apical foramen when the red bars
displayed.A continuous beep sound will be generated at the same time
[Picture 4(c)]
1) Approx 2mm to apical foramen
2) Approx 0.6mm to apical foramen
3) Apex(apical foramen)