21
Attention:
a. When grip the root canal with a needle file, please grip the upper
of the metal part(near the root canal at the needle handle). If you grip the
lower part (bladeor moving part),it will wear the metal part of the file
folder and the resin part. [Picture 7]
b. When measuring the length of root canal, please don’t use the
metal needle file.
If you operate the device without the dentistry glove,it will cause
leakage and the result of measurement will be inaccurate.Therefore,
please use the resin needle file and remember don’t touch the metal part
with finger.
c. Please don’t use the worn file clip, and it will make the result of
measurement inaccurate.
d. Please reference the [Picture 8(a)] to grip the needle file. If as
[Picture 8(b)], it can’t.
4.5.6 When the file reaches the apex, adjust the rubber piece set on
the endo file to the reference point (incision edge or fossa edge), then
pull out the endo file, measure the length between the top of the file and
the rubber piece, and this is the working length of the tooth. It also can
be used with the touch probe instead of file clip, when it is inconvenient
to measure the back teeth. [Picture 9]
4.6 The components that touch body must be autoclaved under
high temperature and high pressure. The shell and measuring wire
should be cleaned by 75% alcohol.
Picture 7 Picture 8(
a
) Picture 8(
b
)