Automatic water analysis system
GENO-softwatch Komfort
Order no. 145 172 955-inter Edited by: phar-mrie G:\BA-172955-INTER_145_GENO-SOFTWATCH KOMFORT
41
Continuation Table G-1: Other errors
This is what you observe
This is the cause
This is what to do
Zero sample error
The set flushing time is
too short.
Prolong the flushing time.
There is no water in the
measuring chamber.
Water pressure too low.
Water inlet is shut off.
The outlet is clogged.
Check whether the hose is
unobstructed.
The plug at the light
source has not been
screwed on properly.
Check the plug connections.
The water pressure is
too low.
Check the water pressure.
There is gas in the sample
(formation of bubbles).
Remedy the introduction of air
into the water.
The measuring chamber
is dirty.
Clean the measuring chamber.
Measuring error
The indicator has been
used up or the indicator
counter has not been
reset after the
replacement of the bottle.
Replace indicator bottle and
reset counter.
The stir bar does not turn
or is missing.
Check whether the stir bar is
turning.
Check whether the stir bar is
in the measuring chamber.
There is no water in the
measuring chamber.
Check the water supply to the
system.
The outlet is clogged.
Check whether the hose is
unobstructed.
There is foreign matter in
the inlet of the solenoid
valve, so that it does not
close completely.
Remove the foreign matter.
Defective hose pump.
Replace hose pump.