Automatic water analysis system
GENO-softwatch Komfort
Order no. 145 172 955-inter Edited by: phar-mrie G:\BA-172955-INTER_145_GENO-SOFTWATCH KOMFORT
32
Note:
In order to operate the rotary switches and to adjust the sliding
switches you will need a small screwdriver. Please only use high-
quality and tested tools in order to avoid damaging these easily
damaged components.
4.1 Factory setting of the rotary switch
Refer to
chapter
Setting
Factory-setting
Pos.
F-4.1.1
Flushing time
4 minutes
6
F-4.1.2
Delay time
Off/0 minutes
0
4.1.1 How to set the flushing time
The flushing time before the start of an analysis can be set to
between 5 seconds and 30 minutes by means of the rotary switch
"flush time" on the left.
Pos. Time
[sec]
Pos. Time
[min]
Pos. Time
[min]
Pos. Time
[min]
0
5
4
1.5
8
10
C
18
1
10
5
2
9
12
D
20
2 20
6 4
A 14
E 25
3
50
7
8
B
16
F
30
Subjecting the flushing time to the length of the feed line to the
GENO-softwatch Komfort ensures that water from the softer is being
analysed.
Note:
The 2 m hose line 6 x 4 with an interior diameter of 4 mm has a
volume of approx. 30 ml and - depending on the pressure conditions -
can be flushed within approx. 4 seconds.
Note:
We recommend a flushing time of at least 20 seconds.