Automatic water analysis system
GENO-softwatch Komfort
Order no. 145 172 955-inter Edited by: phar-mrie G:\BA-172955-INTER_145_GENO-SOFTWATCH KOMFORT
27
2.1 Analysis start – key T1
Manually start an analysis.
When an analysis sequence has been started, you may switch to the
next program step by pressing the start key.
Note:
If an analysis is started manually, the possibly activated relays
REL 1 and REL 2 are deleted.as well.
2.2 Flushing and filling of the measuring chamber – key T2
Outside of an analysis sequence, the measuring chamber and the
feed line to the measuring chamber can be flushed.
2.3 Deaerating dosing pump – key T3
Outside of an analysis sequence, the indicator pump can be switched
on to deaerate the hose line during start-up, for example.
Note:
The stirring device is running together with the indicator pump.
2.4 RESET – key T4
Reset of output relay REL 1 in case the limit value is exceeded (refer
to analysis result, chapter F-3).
Reset of output relay REL 2 in case of a device malfunction or the
signal "lack of indicator" (refer to alarm signals, limit value exceeded
chapter F-3.4).
Note:
By means of these keys, an analysis in progress can be
aborted
and previous alarms can be deleted.
2.5 Resetting the filling volume for the indicator – keys T3 and T4
By simultaneously pressing the keys, the measurement of the
indicator volume is reset. The reset has been carried out if the LEDs
lack of indicator (blue) and alarm signal (red) light up at the same
time.
Note:
The reset is only allowed to take place if a full 500 ml indicator
bottle has been inserted.
T1
T2
T3
T4
T3
T4