Automatic water analysis system
GENO-softwatch Komfort
Order no. 145 172 955-inter Edited by: phar-mrie G:\BA-172955-INTER_145_GENO-SOFTWATCH KOMFORT
36
4.2.3 Function relay REL 1
Relay REL 1 (terminal 9/10/11) signals that the limit value has been
exceeded. You choose between a pulse contact of 3 and 60 seconds
for the activation of a controller or a permanent contact.
Regarding a permanent contact, two alternatives are available:
a. Analyses are performed continuously at the set time interval. If the
value falls below the limit value, the relay REL 1 has dropped out.
b. No further analyses will be performed after the limit value has
been exceeded. Relay REL 1 must be reset by pressing the
RESET key. Only afterwards analyses will be carried out again.
Function of relay REL 1
S4 S5
Function
OFF OFF
Pulse contact
3 seconds
OFF ON
Pulse contact
60 seconds
ON OFF
Permanent contact
No analysis stop
(refer to a)
ON ON
Permanent contact
Analysis stop
(refer to b)