50
GMC-I Messtechnik GmbH
6.3
Uploading a Report Form
Report and print-out forms can be generated and modified with
the help of WinProfi software, and can be uploaded to the
PROFITEST 204+
. WinProfi software provides the user with a high
performance editor to this end.
Note
Before uploading a report form, the header should be
edited to fulfill your needs (see chapter 6.4.2, page 54).
GMC-I Messtechnik GmbH’s name and address are
included in the header upon shipment from the factory.
!
Attention!
All measurement data stored to the test instrument are
deleted when a print form is uploaded!
There are two recommended procedures for uploading report
forms:
• 1. Upload the report form.
2. Perform measurements.
3. Print the report.
or
• 1. Perform measurements.
2. Transmit and save measured values to a PC.
3. Upload the report form.
4. Transmit measured values back to the
PROFITEST 204+
.
5. Print the report.
A Installing and Starting WinProfi at the PC
➭
Install the software as described in chapter 3.7.5, part A, if not
already installed.
➭
Start WinProfi from the Windows start menu.
B Data Transmission Prerequisites
➭
Establish a connection between the PC and the test
instrument. Use the Z3241 cable to this end which is included
with the test instrument.
➭
Switch both devices on.