40
GMC-I Messtechnik GmbH
Storing Measured Values to Memory
After the test sequence has been completed, the last measured
values for U
T
and I
T
remain at the display. These results can be
saved with or without a comment, and a consecutive number is
assigned to the test automatically.
Up to 15 characters can be entered to the entry field, or to the info
field at the bottom left as a description for the measurement. See
chapter 5.1.1, page 43, regarding data entry.
The values at the display are overwritten when the next test is
executed.
STORE
Press briefly:
Save results.
Press and hold:
Store an event and
activate the entry
field.
No values are saved to memory for U
T
and I
T
during continuous
operation (with test duration set to
>>>>>>>
).
Premature Interruption of the Test
The test can be interrupted prematurely at any time by releasing
the trigger at the high-voltage pistol with switch.
If testing is interrupted by releasing the trigger at the high-voltage
pistol before the pre-selected test duration has elapsed, the val-
ues displayed for U
T
and I
T
may deviate from the last measured
values.
Note
Regarding Breakdown Voltage:
If the selected breaking current I
MAX
is
exceeded before the selected test voltage
value has been reached, the test voltage
measured at this point in time is preceded
with the “>” symbol at the display, and is stored to
memory. The corresponding symbol appears as well.